• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 23AGFN-756
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Insufficient Information (4580)
Event Date 07/14/2022
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It as reported that on (b)(6) 2022, a 23mm sjm regent heart valve with flex cuff was selected for procedure.It was noted that during procedure a leaflet of the regent valve become dislodged.The decision was made to replace the 23mm sjm regent heart valve with flex cuff with another one of the same size.However, it was noted that there was a 60 minute delay in procedure because the replacement valve had to be delivered from a different hospital.The patient was discharged at the time of report.
 
Manufacturer Narrative
An event of a dislodged leaflet was confirmed.One leaflet was unfractured and dislodged from the orifice.No other damages were noted on the valve.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The root cause of the leaflet dislodgement could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15298682
MDR Text Key298666873
Report Number2135147-2022-00916
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734006095
UDI-Public05414734006095
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number23AGFN-756
Device Catalogue Number23AGFN-756
Device Lot Number7624233
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/31/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/13/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age54 YR
Patient SexMale
-
-