LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-80 |
Device Problem
Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/27/2022 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report of a heater-cooler system 3t power failure.Reportedly, the device could not be powered on.There was no patient involvement.
|
|
Manufacturer Narrative
|
The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The failure was traced back to a defective ac mains board caused by a short circuit of the cardioplegia tank heating element.The heating element and the mains board are going to be replaced on a return visit to complete the repair.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
H.10: the customer decided to not repair the device due to its age and replaced it with a new one.However, the root cause of the reported event can be traced back to an electrical failure of the cardioplegia heating element.
|
|
Event Description
|
See initial report.
|
|