• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/27/2022
Event Type  malfunction  
Event Description
Livanova deutschland received a report of a heater-cooler system 3t power failure.Reportedly, the device could not be powered on.There was no patient involvement.
 
Manufacturer Narrative
The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The failure was traced back to a defective ac mains board caused by a short circuit of the cardioplegia tank heating element.The heating element and the mains board are going to be replaced on a return visit to complete the repair.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
H.10: the customer decided to not repair the device due to its age and replaced it with a new one.However, the root cause of the reported event can be traced back to an electrical failure of the cardioplegia heating element.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key15299784
MDR Text Key304697333
Report Number9611109-2022-00429
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/07/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-