• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/20/2022
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova deutschland manufactures the s5 double head pump.The incident occurred in homewood, alabama.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The motor control board and motor driver board.The problem could be solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova received report that a s5 system displayed an error message "no monitoring/control pump".A s5 double head pump stopped, gave an audible beeping and touchscreen was not displaying time/volume delivered.The perfusionist power cycled the pump but this did not correct the problem.Perfusionist swapped the device with a spare unit and successfully completed the case.There was no patient injury.
 
Manufacturer Narrative
H.10: based on service activity findings, the root cause of the reported issue can be traced back to faulty motor controller board and motor driver board.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key15299785
MDR Text Key300246958
Report Number9611109-2022-00432
Device Sequence Number1
Product Code DWB
UDI-Device Identifier04033817900405
UDI-Public(01)04033817900405(11)180416
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/16/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-