Model Number 10-85-00 |
Device Problem
Pumping Stopped (1503)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/28/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova deutschland manufactures the s5 double head pump.The incident occurred in milano, italy.A livanova field service representative was dispatched to the facility and replaced the motor control board as precautionary measure.Subsequent functional verification testing was completed without further issues.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova received report that a s5 double roller pump gave a motor control failure error message during priming.There was no patient involvement.
|
|
Manufacturer Narrative
|
H.10: the serial read-out of the pump (real time device parameters and setting recording file) was analyzed and confirmed that on the date of the event the reported error code was stored in the pump microcontroller.The reported error code can be solely related to a fault of the motor control board of the pump which was replaced as per initial report.Based on collected information and service findings, the root cause of the reported event is traced back to an intermittent malfunction of the motor control board.
|
|
Event Description
|
See initial report.
|
|