Model Number EG29-I10 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/22/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
If additional information becomes available, a supplemental report will be filed with the new information.
|
|
Event Description
|
Customer reported frozen image.This event occurred at the time of before use.There was no report of patient harm.
|
|
Manufacturer Narrative
|
Correction information: follow up #1.If follow-up, what type? device evaluated by manufacturer.Adverse event problem: coding changed based on the investigation result.Additional information device manufacture date.Evaluation summary as a result of the investigation, it is determined that the ccd and cable are out of order due to external load.
|
|
Manufacturer Narrative
|
Correction information g2: report source (checked "foreign", so fixed it.) g6: follow up #2 h2:if follow-up, what type?.
|
|