• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. FAN SPRAY KIT; DISPOSABLE BATTERY-POWERED SURGICAL/ORTHOPAEDIC LAVAGE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. FAN SPRAY KIT; DISPOSABLE BATTERY-POWERED SURGICAL/ORTHOPAEDIC LAVAGE SYSTEM Back to Search Results
Model Number N/A
Device Problems Failure to Power Up (1476); Battery Problem (2885); Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2022
Event Type  malfunction  
Event Description
It was reported that the device didn't work at all, the battery pack heated up and there was a battery leakage.The event timing was during surgery.There was no harm and delay was less than 15 minutes.An alternate device was available for use.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.
 
Manufacturer Narrative
Functional testing found the device would not power on in high or low mode with the original battery pack but did power on in both modes when connected to a lab battery pack.Visual inspection if the interior of the pack found the batteries had leaked electrolyte inside of the pack.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
DISPOSABLE BATTERY-POWERED SURGICAL/ORTHOPAEDIC LAVAGE SYSTEM
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15310822
MDR Text Key303326467
Report Number0001526350-2022-00842
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00889024375161
UDI-Public(01)00889024375161(17)250130(10)65390963
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00515047501
Device Lot Number65390963
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-