SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX CONTINUOUS EPIDURAL TRAY; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 4935MP-17
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2020
Event Type  malfunction  
Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found the first photo shows a fragment of the tray lid and the second photo shows the tray with two components.One of the components appeared to be compromised.Evaluation of the provided samples with unaided eye found two devices, one of the devices was observed to be broken and the wire was stretched.Review of the bill of material for tray part and the set up picture, included in device history record (dhr) found no components that could be matched with the returned samples.These samples do not resemble the needle nor the catheter from the affected tray.The provided samples are not provided as a part of the tray.The root cause of the reported issue was found to be not relevant to this manufacturer medical product.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
 
Event Description
It was reported a continuous epidural that staff stated unwound.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX CONTINUOUS EPIDURAL TRAY
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
olomoucka 306
hranice 1, mesto 753 0 1
EZ  753 01
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
olomoucka 306
hranice 1, mesto 753 0 1
EZ   753 01
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15311632
MDR Text Key305355021
Report Number3012307300-2022-16523
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K965017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2021
Device Catalogue Number4935MP-17
Device Lot Number4009465
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-