NULL PORTEX CONTINUOUS EPIDURAL ANESTHESIA TRAY; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 4509E
Device Problem Break (1069)
Patient Problem Urethral Stenosis/Stricture (4501)
Event Date 09/01/2020
Event Type  malfunction  
Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual testing was performed.Visual inspection found the hinge part of all the connectors were damaged.The root cause of the reported issue was found to be the supplier.An internal scar has been opened.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional corrective actions will be taken.Udi, operator of device, and mfr site are unknown.No information has been provided to date this remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that immediately after starting to use the product, the customer noticed the connection part of the epidural connector was broken.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX CONTINUOUS EPIDURAL ANESTHESIA TRAY
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section G)
NULL
MDR Report Key15315629
MDR Text Key305251721
Report Number3012307300-2022-16586
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K781052
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number4509E
Device Lot Number200522
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/11/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-