AVANOS MEDICAL INC. BALLARD CLOSED SUCTION SYSTEMS FOR NEONATES/PEDIATRICS, ELBOW WITH CAPS; VAP CLOSED SUCTION CATHETERS & ACCESSORIES
|
Back to Search Results |
|
Model Number 20083 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/02/2022 |
Event Type
malfunction
|
Event Description
|
It was reported, the control valve on the closed suction catheter (csc) was verified to be sticking; the issue was discovered during use.There was no reported injury.Additional information received 04aug2022 reported, the thumb valve was not working properly; the csc was in place for ¿about 1 hour; there was no reported injury.
|
|
Manufacturer Narrative
|
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 29 aug 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|
Manufacturer Narrative
|
Visual examination of the device: the control valve was examined inside and did not appear to have visible damages, there was a slight scratch on the side of the cap and the position of the thumb valve appeared slightly in a downward position.Testing of the sample: the thumb valve was rotated to the locked position with no issue, when the valve was pressed and released it did not return to the full upright position; however, the valve could be manually pulled up into the full upright position.Suctioning with the device failed (continual suctioning was observed) due to the thumb valve being unable to return to the complete upright position.The reported event could be confirmed as reported.The root cause is traced to the manufacturing process; preventative measures are in place.The device history record for lot 30161904 was reviewed and the product was produced according to product specifications.All information reasonably known as of 11 oct 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|
Search Alerts/Recalls
|
|
|