• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL BCI SPECTRO2 30 PULSE OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL BCI SPECTRO2 30 PULSE OXIMETER Back to Search Results
Catalog Number WW1030DE
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned for evaluation.Visual and functional testing were performed.The device looked to be in new condition, no cracks or damage visible.Functional testing duplicated the reported problem.The root cause of the reported issue was found to be a cracked light-emitting diode (led) caused by plastic light too close to red led.The led was replaced.The device history record (dhr) review was not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.No product information has been provided to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
 
Event Description
It was reported that the device red light-emitting diode (led) alarm was not working.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI SPECTRO2 30 PULSE OXIMETER
Type of Device
OXIMETER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15322836
MDR Text Key305357429
Report Number3012307300-2022-16847
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K083705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberWW1030DE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2021
Was the Report Sent to FDA? No
Date Manufacturer Received05/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-