U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES¿ ULTRA 3D IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES¿ ULTRA 3D IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1601-04
Device Problems Mechanical Problem (1384); Energy Output Problem (1431)
Patient Problems Failure of Implant (1924); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 09/16/2022
Event Type  malfunction  
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Event Description
The recipient is reportedly experiencing loudness growth issues.Revision surgery is scheduled.
 
Manufacturer Narrative
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Dvanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed cuts in the silicone overmold on the top and bottom covers, as well as a severed electrode.These are believed to have occurred during revision surgery.Photographic imaging inspection revealed a broken electrode within the electrode pocket.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the hires ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
heidi white
28515 westinghouse place
valencia, CA 91355
MDR Report Key15324352
MDR Text Key300660386
Report Number3006556115-2022-01345
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016862250
UDI-Public(01)07630016862250(11)190927(17)220831
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/31/2022
Device Model NumberCI-1601-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age19 YR
Patient SexFemale
-
-