ST PAUL BCI SPECTRO2 30 PULSE OXIMETER Back to Search Results
Model Number WW1020EN
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This monitor returned to service about 2 years later with missing holding pads causing the spo2 board to loose.With the tamper seal not applied to the monitor at the time of manufacturing, unable to determine whether these missed during manufacturing or from customer tampering with the unit.The dhr review is not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.Operator of device is unknown.No information provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that the device is showing sr error.No patient injury has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI SPECTRO2 30 PULSE OXIMETER
Type of Device
OXIMETER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15326717
MDR Text Key301092302
Report Number3012307300-2022-16932
Device Sequence Number1
Product Code DQA
UDI-Device Identifier30843418000108
UDI-Public30843418000108
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberWW1020EN
Device Catalogue NumberWW1020EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-