Model Number WW1030EN
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
First time in for service after 2 years in the field and the reported problem is a user perception resulting from user interfacing with the unit in a manner inconsistent with the ifu.It is a normal operation of the spectro2 to alert the customer when no reading is being acquired or no patient is being detected because a patient is not attached.The dhr review is not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.No information provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.
 
Event Description
It was reported that the device is giving an error led that lights up.No patient injury has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI SPECTRO2 30 PULSE OXIMETER
Type of Device
OXIMETER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15326757
MDR Text Key305368816
Report Number3012307300-2022-16935
Device Sequence Number1
Product Code DQA
UDI-Device Identifier30843418000191
UDI-Public30843418000191
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083557
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberWW1030EN
Device Catalogue NumberWW1030EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/13/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-