The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
|