It was reported that 4 months after placement of dxdt2212, in-growth occurred.Therefore, another stent was additionally placed.7 months later, both stents were found migrated.It was confirmed from the device history record that device had been manufactured with no significant issues and passed all the inspections successfully.Duodenum structure where stent was implanted is curvy.It is possible that the stent could be pressed and stent in-growth could occur by state of patient's lesion.In addition, migration can occur in any company's device as well as ours.It is affected by patient's lesion status, peristalsis of organs, and drug use in general.However, it is hard to identify the exact root cause since it is hard to reconstruct the situation at the time of procedure.It is difficult to identify the exact root cause because the device was not returned, and it is difficult to reconstruct the situation at the time of procedure.However, based on the description "another stent was additionally placed in stent-in-stent method due to in-growth" and "recovering the stenosis thanks to the chemotherapy could be a factor of stent migration", it is assumed that in-growth occurred on the dxdt2212 which is an uncovered stent due to the pressure of patient's lesion etc.Then after placement of another stent, stent migration occurred due to chemotherapy, condition of the patient's lesion, peristalsis of organs, foreign substances such as food and other factors complexly.In addition, there is a possibility the patient felt difficult to eat due to the stent migration, therefore it is assumed that dxdt2212 was additionally placed.Through the user manual by taewoong, it is stated that "potential complications associated with the use of niti-s & comvi stent may include, but are not limited to: stent migration".This suspected device is not registered in the us but we will continuously monitor the same or similar customer complaints through accurate analyses.
|