Catalog Number PTB086275W
Device Problems Separation Failure (2547); Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2022
Event Type  malfunction  
Manufacturer Narrative
Patient weight was unavailable.A review of the manufacturing records indicated the device met pre-release specifications.Neither clinical images enabling assessment of product performance was returned for evaluation nor the product itself (remains implanted) was accessible for testing, the investigation could not be performed.Cause of the event cannot be established based on the available information.Further information regarding this event was requested by gore, but no further information has been reported, therefore, this investigation is considered complete.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
The following was reported to gore: on (b)(6) 2022, a patient was to be implanted with a 8mm x 6cm gore® tips viatorr® (viatorr® device) during a tips procedure.The viatorr® device was advanced to the target lesion.The unlined region of device expanded successfully.The deployment knob was pulled, it was found that the viatorr® device couldn't complete the expansion at around 0.5-1cm away from head end (proximal end) and the deployment line stuck.Physician tried to twist the delivery catheter, but the device still couldn't complete the expansion.A balloon was used to dilate the device to 26 atmospheres, trying to expand the device, the deployment line showed a little loose but the device couldn't fully expand by pulling the deployment line.Due to there was no good solution after consulting physicians from different departments, the physician decided to cut the deployment line that was outside patient and have other parts left in patient with one end tied to device.Patient tolerated the procedure, situation was stable and discharged from hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIATORR® TIPS ENDOPROSTHESIS
Type of Device
SHUNT, PORTOSYSTEMIC, ENDOPROSTHESIS
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 1 B/P
32360 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
qiong wang
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15367936
MDR Text Key305982087
Report Number3007284313-2022-02101
Device Sequence Number1
Product Code MIR
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P040027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPTB086275W
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/26/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
-
-