|
Catalog Number VS-403 |
Device Problems
Biocompatibility (2886); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Inflammation (1932); Pain (1994); Phlebitis (2004)
|
Event Date 08/06/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Physician used venaseal occluding device during procedure to treat the great saphenous vein (gsv).General anaesthesia was used.Hand was used for compression.The lumen was flushed prior to use.Ifu was followed.A guidewire was used for the insertion of the catheter.It was reported that a day post procedure, patient presented with irritation near the access point.Irritation is still present.Physician used closurefast for above the knee and venaseal for below the knee.Skin irritation was observed along venaseal treated segment.Other symptoms in addition to skin irritation were pain and tenderness along the treated segment.Physician had complained of similar events with venaseal new version vs-403 and suspect something is wrong with the new venaseal.Complications associated with this event is phlebitis.Only one leg, using vs-403.Previous leg was treated a year ago with sp-101.The catheter tip was 5cmcaudal to sfj.Steroids were prescribed by physician.No patient injury reported.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|