It was reported that a 25-week-old baby was found to have sepsis following a leak that occurred during an infusion involving a 25 cm (10") pur yellow smallbore ext set, 6-port nanoclave¿ manifold w/check valve, nanoclave¿, 4-way nanoclave¿ stopcock (red ring), rotating luer in the pediatric intensive care unit.The leak was reported to have occurred at the level of the small yellow extension while infusing the following medications: parenteral nutrition by numetah g13, insulin, morphine, primene.The clinician noticed that the sheet was wet and that the liquid was coming out at the end of the 6-port manifold.The leak was reported to be in contact with the patient.The sheet was reported to have been wet and subsequently changed.In addition, the central catheter line was replaced according to hospital service protocols.The reporting clinician is unable to attribute the patient¿s sepsis to this incident.There was a delay in therapy; however, no additional medical intervention was reported.No additional information is available at this time.
|
Received one (1) used 011-am6136 pur yellow smallbore ext set and one (1) new 011-am6136 pur yellow smallbore ext set for inspection.As received, the collar on the male luer was freely spinning on the used set.The sets were tested according to product specifications.On the used set, there was a channel leak at the connection of the male luer to the female luer of the yellow tubing.The male collar was cut and the bonded connection of the male luer and female luer of the yellow tubing was disassembled.There was no leakage on the new set.The new set met all design specifications.The reported complaint can be confirmed on the received one (1) used 011-am6136 pur yellow smallbore ext set.The probable cause of the channel leak is due to an incomplete bond during manufacturing.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.Additional information: product receipt date 10/13/2022.
|