Catalog Number 110003452 |
Device Problems
Positioning Failure (1158); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/16/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the modular cup inserter threaded adapter would not screw into the cup and the threads on the shaft seem to be worn down.No adverse consequences were reported.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-02071.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
No additional information on the reported event.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified damage to the 1st thread of the inserter.The thread is deformed.Nicks and dings are present on the locking feature.The surface of the shaft is scratched.The hex feature at each end of the inserter is scratched and gouged.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|