Model Number 71422425
Device Problem Failure to Osseointegrate (1863)
Patient Problem Inadequate Osseointegration (2646)
Event Date 09/06/2021
Event Type  Injury  
Event Description
It was reported that after a left tka performed on (b)(6) 2021 due to osteoarthritis, the patient experienced aseptic loosening of the tibial component.This was addressed via revision surgery on (b)(6) 2021 were all components where exchange.This information was provided by the national joint registry of the united kingdom, as part of a retrospective data collection of patients who underwent a revision surgery of journey uni prothesis.As such, no further information will be available.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that without the requested medical documentation, possible clinical factors which could have contributed to the reported event could not be definitively concluded.The patient impact beyond that which was reported could not be determined.No further medical assessment can be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that loosening event has been identified in possible adverse effects section as a result from trauma, improper implant selection, improper implant positioning, improper fixation, and/or migration of the components.Muscle and fibrous tissue laxity can also contribute to these conditions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include abnormal motion over time, bone degeneration, size selected, inadequate integration between the cement and bone/implant, osteolysis and/or traumatic injury.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: case-2022-00119086-1 section d10 was corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY UNI TIBIAL BASE LM/RL SZ 5
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15385563
MDR Text Key299552218
Report Number1020279-2022-04030
Device Sequence Number1
Product Code HSX
UDI-Device Identifier00885556088043
UDI-Public00885556088043
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K073175
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71422425
Device Catalogue Number71422425
Device Lot Number20MM01053
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Treatment
71422261 JOURNEY UNI TIBINRT S5-6LM/RL8MM.; 71422425 JRNY UNI TIBIAL BASE LM/RL SZ 5.; JOURNEY UNI TIBINRT S5-6LM/RL8MM, LOT: 20KM14412.; JUNI OX FB FEM SZ 5 LM RL, LOT: 19MM04028.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexMale
-
-