Related manufacturer reference number: 3006705815-2022-16418, 1627487-2022-04900.It was reported that during ipg explant procedure on (b)(6) 2022 (related manufacturer reference number: 3006705815-2022-16418), the leads were cut and left in situ.Surgical intervention may be undertaken to address this issue.
|
Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
|