Model Number MX9505T
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/15/2020
Event Type  malfunction  
Event Description
It was reported that liquid leakage was found during use.No patient injury was reported.
 
Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Two product sample were received for evaluation.Visual and functional testing were performed.It was observed that a component from a subassembly had lack of solvent on the tube.One sample from the two was found leaking on the subassembly when doing a water leak test.Relevant documents were reviewed and deemed adequate and correct with respect to investigation activities.The complaint is confirmed.The most likely root cause is related to a supplier issue.Operator of device unknown.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSTAR PRESSURE TRANSDUCER
Type of Device
PROBE, BLOOD-FLOW, EXTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6250 shier rings road
dublin OH 43016
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
6250 shier rings road
dublin OH 43016
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15397881
MDR Text Key300595643
Report Number3012307300-2022-18562
Device Sequence Number1
Product Code DPT
UDI-Device Identifier10351688503992
UDI-Public10351688503992
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K942377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date06/14/2022
Device Model NumberMX9505T
Device Catalogue NumberMX9505T
Device Lot Number3833910
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2020
Was the Report Sent to FDA? No
Date Manufacturer Received05/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-