• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER; SIMILAR DEVICE D133601, PMA # P030031/S053

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER; SIMILAR DEVICE D133601, PMA # P030031/S053 Back to Search Results
Catalog Number D133604IL
Device Problems Material Puncture/Hole (1504); Material Twisted/Bent (2981); Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/15/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone: (b)(6).The product investigation was completed.Device evaluation details: visual analysis of the returned sample revealed a hole was found in the pebax, and the tip was found bent; no other damage or anomalies were observed on the smart touch unidirectional catheter.The damage could be related to excessive force or to the shipment process a deflection test was performed, and the curve was deflecting within specifications.No deflection issues were observed.The deflection issue reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the usage of the device that compromised its performance.A manufacturing record evaluation was performed for the finished device number 30699294m, and no internal actions related to the complaint were found during the review.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Event Description
A patient underwent an ablation procedure with a thermocool® smart touch¿ electrophysiology catheter for which biosense webster¿s product analysis lab identified a hole on the pebax and the tip bent.It was initially reported that during the procedure, the catheter was unable to deflect or relax completely.A second catheter was used to complete the operation.There was no adverse event reported on the patient.Deflection issue is not mdr-reportable.Bent tip is not mdr-reportable.Hole on the pebax is mdr-reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER
Type of Device
SIMILAR DEVICE D133601, PMA # P030031/S053
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-798
MDR Report Key15398120
MDR Text Key299964927
Report Number2029046-2022-02196
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 09/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/09/2023
Device Catalogue NumberD133604IL
Device Lot Number30699294M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2022
Date Manufacturer Received08/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-