Catalog Number V041163000000 |
Device Problems
Mechanical Problem (1384); Failure to Power Up (1476); Output Problem (3005)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
In this event it is reported that a vdw silver reciproc stops during use in treatment or does not turn on.Outcome of patient injury is pending.
|
|
Manufacturer Narrative
|
There has been a previous report received where this malfunction has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Manufacturer Narrative
|
Additional information was received indicating that no injury resulted.Battery (voltage collapses under load) defective, motor connection cable (torn strain relief) defective, foot control 103543 (loose contact) defective, micromotor smr16409 tested, without errors.Supplied without contra angle and power supply.
|
|