The customer reported to olympus, during a diagnostic esophagogastroduodenoscopy, the loop cutter became stuck after cutting through a suture.The procedure was completed using an unknown device to remove the loop cutter.The customer reported the device was inspected prior to the procedure with no anomalies found.There was no harm or injury to the patient associated with this event.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.Based on the results of the investigation, although the reported event could not be duplicated, it is likely that the issue occurred because the user attempted to cut sutures with the product that was outside the intended use.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿before each case, prepare and inspect the instrument as instructed below.Inspect other equipment to be used with the instrument as instructed in their respective instruction manuals.Should the slightest irregularity be suspected, do not use the instrument; contact olympus.Damage or irregularity may compromise patient or user safety, such as punctures, hemorrhages or mucous membrane damage and may result in more severe equipment damage.¿ ¿do not try to cut the loop that is not positioned on both edges of the loop hanger as plumb as possible for the blade.It may make cutting the loop impossible or result in the loop getting caught in the distal end of the instrument, which could make it difficult or impossible to remove from the patient.In this case, use pliers to cut the insertion portion of the instrument where it extends from the biopsy valve of the endoscope.Remove the endoscope from the body, then reinsert the endoscope and cut the loop with a spare loop cutter.¿ ¿do not use the instrument to cut any objects other than the surplus of olympus loop (e.G., maj-254, maj-340).If anything, other than the surplus of the loop is cut, the blades may be damaged and unable to perform properly, the cut object may be caught in the tip of the instrument, and it may become difficult to safety remove the instrument from the body.Such objects other than the surplus of loop may include, stent wire, sewing threads, and loop stoppers.¿ ¿do not try to cut the loop that is not positioned on both edges of the loop hanger as plumb as possible for the blade.It may make cutting the loop impossible or result in the loop getting caught in the distal end of the instrument, which could make it difficult or impossible to remove from the patient.In this case, use pliers to cut the insertion portion of the instrument where it extends from the biopsy valve of the endoscope.Remove the endoscope from the body, then reinsert the endoscope and cut the loop with a spare loop cutter.Do not use the loop cutter to cut anything other than the loop.If you cut any other object, it may get caught in the distal end.This could make it difficult or impossible to remove it from the patient.¿ ¿do not cut the loop unless you have a clear endoscopic field of view.This could cause patient injury, such as punctures, hemorrhages or mucous membrane damage.It may also damage the endoscope and/or instrument.¿ olympus will continue to monitor field performance for this device.
|