NULL PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Lot Number 200603 |
Device Problems
Device Alarm System (1012); Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/01/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during a hospital pre-use check, the anesthesia device issued an air leak alarm, and would not allow the customer to use the product.No patient injury was reported.
|
|
Manufacturer Narrative
|
Two sample kits were received to perform an investigation.A device history record (dhr) review was conducted for the received devices which indicated all inspections were completed and no issues were noted during manufacture.A visual inspection found no breakage or deformation of the returned samples.Functional testing was performed using a leak test.No leakage of the anesthesia circuit, heat and moister exchanger (hme), or anesthesia bag was found.All devices conformed to the product specifications.No root cause could be determined as the complaint could not be confirmed.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
|