• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NULL PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NULL PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Lot Number 200603
Device Problems Device Alarm System (1012); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2020
Event Type  malfunction  
Event Description
It was reported that during a hospital pre-use check, the anesthesia device issued an air leak alarm, and would not allow the customer to use the product.No patient injury was reported.
 
Manufacturer Narrative
Two sample kits were received to perform an investigation.A device history record (dhr) review was conducted for the received devices which indicated all inspections were completed and no issues were noted during manufacture.A visual inspection found no breakage or deformation of the returned samples.Functional testing was performed using a leak test.No leakage of the anesthesia circuit, heat and moister exchanger (hme), or anesthesia bag was found.All devices conformed to the product specifications.No root cause could be determined as the complaint could not be confirmed.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section G)
NULL
MDR Report Key15410202
MDR Text Key305591905
Report Number3012307300-2022-18899
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Lot Number200603
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2020
Was the Report Sent to FDA? No
Date Manufacturer Received07/08/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-