Model Number N/A |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/08/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Foreign country: taiwan.The event is confirmed.Review of the most recent repair record determined the motor speeds were out of specification on the low end.The motor, reciprocating arm, o-ring, and various bearings were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that the device was in need of repair due to a connection problem.The motor, bearing pack, o-ring, seal, and reciprocating arm were defective.The device was not a new device.The event occurred during a routine inspection of the device.The investigation found that the motor speeds were out of specification on the low end.No adverse events were reported as a result of this malfunction.Attempts have been made and no further information has been provided.
|
|