Model Number B1050-020 |
Device Problems
Inflation Problem (1310); Material Rupture (1546)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/24/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that the procedure was to treat an unspecified lesion.A 5x20mm armada 35 balloon was advanced to the lesion without issue and when inflated once at nominal pressure, a leak was noted in the balloon catheter area and the balloon failed to inflate.Another armada 35 was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
Visual and functional inspections were performed on the returned device.The reported balloon rupture and inflation issue were not confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported complaints.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
|
|