• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. CONSULT HCG DIPSTICK TEST 5000 25T; PREGNANCY TEST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. CONSULT HCG DIPSTICK TEST 5000 25T; PREGNANCY TEST Back to Search Results
Model Number FHC-101
Device Problem False Positive Result (1227)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Pending investigation results.This mdr is to document false positives observed at location 2 out of 3.See related mdr's 2027969-2022-00074 and 2027969-2022-00076 for location 1 and 3.
 
Event Description
Unspecified date: four confirmed false positive results on the consult hcg dipstick kit.Unknown number of patients involved, however the false positive results occurred across three different locations.Samples were sent to quest for confirmatory testing and a negative result was obtained; quantitative result not provided.No adverse outcomes reported.Although requested, no further information was provided.This mdr is to document false positives observed at location 2 out of 3.See related mdr's 2027969-2022-00074 and 2027969-2022-00076 for location 1 and 3.
 
Manufacturer Narrative
Update to d9: device available for evaluation- no.H3: although requested, the device was not returned for investigation.Retained devices from the reported lot number were tested with hcg-negative clinical urine samples.The results were read at 3 and 4 minutes and all devices yielded the expected negative results.No false positive results were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention product.Complaints are tracked and trended on a monthly basis.Per the package insert: very low levels of hcg (less than 50 miu/ml) are present in urine specimen shortly after implantation.However, because a significant number of first trimester pregnancies terminate for natural reasons, a test result that is weakly positive should be confirmed by retesting with a first morning urine specimen collected 48 hours later.A number of conditions other than pregnancy, including trophoblastic disease and certain non-trophoblastic neoplasms including testicular tumors, prostate cancer, breast cancer, and lung cancer, cause elevated levels of hcg.Therefore, the presence of hcg in urine specimen should not be used to diagnose pregnancy unless these conditions have been ruled out.This test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy should only be made by a physician after all clinical and laboratory findings have been evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSULT HCG DIPSTICK TEST 5000 25T
Type of Device
PREGNANCY TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge rd.
san diego CA 92121
Manufacturer Contact
monika burrell
9975 summers ridge rd.
san diego, CA 92121
8588052506
MDR Report Key15414951
MDR Text Key305796550
Report Number2027969-2022-00075
Device Sequence Number1
Product Code JHI
UDI-Device Identifier20612479202751
UDI-Public(01)20612479202751(17)240131(10)HCG2022003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993203
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Model NumberFHC-101
Device Lot NumberHCG2022003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/07/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/19/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-