NULL PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C49101320J |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/01/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that there was no anomaly in the product at a pre-use check.However, when using the product during the operation, an alarm suddenly went off.The customer checked the product and noticed the corrugated tube was considerably torn.No patient injury was reported.
|
|
Manufacturer Narrative
|
No lot number was provided; therefore, device history record review could not be completed.A product sample and pictures were received for evaluation.Visual and functional testing were performed.Pictures received it is observed that the circuit is crushed.The samples were visually inspected at a distance of 12 under normal lighting to received unit, in order to detect any damage on the units.The circuits have a tear.The reported issue was confirmed.A corrective and preventative action has been opened in order to determine the root cause and implement proper corrective actions for leaking.D4 udi is unknown.D5 operator of device is unknown.G2 manufacturing site address is unknown.G5 pma/510k is unknown.No information has been provided to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
|
|