Model Number ROB10015
Device Problems Connection Problem (2900); Unintended Application Program Shut Down (4032)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that when hitting unlock to insert the bur, a communication failure screen popped up.After hitting ok, an internal error screen popped up.They exited the case, took parts apart from the drill, unplugged the drill, entered the case, plugged the drill back in, put the drill array and long attachment onto the drill ensuring the long attachment was locked, hitting unlock to insert the bur; but the failure happened again.As this was noticed during set up, the cori assisted tka surgery was performed, without any delay, using a ri robotic drill attachment back up.Patient was not impacted.
 
Manufacturer Narrative
H3, h6: the real intelligence robotic drill attachment, part number rob10015, serial number (b)(6), intended for use in treatment, was returned for evaluation.The reported problem could not be confirmed with a visual inspection.The reported problem was not confirmed with a functional evaluation.The drill attachment was able to be attached to a test lab drill and all diagnostics passed multiple times and multiple test case setups were completed.A relationship between the reported event and the device could not be established.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with a drill attachment retaining nut coming loose due to vibration during use.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The failure mode and associated risk have been anticipated within the risk file.The risk level is still adequate.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RI ROBOTIC DRILL ATTACHMENT
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15415438
MDR Text Key304697455
Report Number3010266064-2022-00602
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757345
UDI-Public00885556757345
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10015
Device Catalogue NumberROB10015
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
REAL INTELLIGENCE CORI, SN:(B)(6).
-
-