Device analysis: the returned product consisted of a watchman flx delivery system with the implant in the sheath, and blood in the hub of the device.The implant, core wire, tip, sheath, and hub/valve were microscopically and visually examined.Inspection revealed that there were numerous kinks throughout the sheath and there was tip damage as the tri-cuts were flared and there were abrasions on the inside of the sheath at the tip.Functional testing was completed by attaching a syringe filled with water, and positive/negative pressure was applied with the valve open and closed.The device could not be flushed properly.The device was not able to backflush, and air could not be purged from the device, so the valve was removed and microscopically examined.The silicone valve was found to be damaged.There was no other damage or defect found on the remainder of the device.Product analysis confirmed the reported event, as the silicone valve was damaged, causing difficulty with the flushing of the device.
|