Model Number 10-80-00 |
Device Problem
Pumping Stopped (1503)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/28/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova deutschland manufactures the s5 roller pump.The incident occurred in cotignola, italy.A livanova field service representative was dispatched to the facility to investigate the device.The motor control board and the motor end stage board were replaced due to oxidized parts caused by fluid fallen on the circuit boards.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova received report that a s5 roller pump stopped during procedure and an error was displayed.After restart of the pump the error did not re-appear.There was no patient injury.
|
|
Manufacturer Narrative
|
H.10: pump serial read-out analysis was performed and revealed an error code associated to motor not connected stored in the micro-controller which confirmed the reported event.Moreover, a runaway_peak message was stored as well indicating that the pump was hand cranked without having been switched off.Complaints database analysis revealed no similar event since unit installation in 2008.Based on available information, the root cause of the event can be traced back to motor control board and the motor end stage board components oxidation.
|
|
Event Description
|
See initial report.
|
|