• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. AZURION; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. AZURION; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number AZURION 7 M20
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2022
Event Type  malfunction  
Event Description
It has been reported to philips that during a planned treatment, the l-arm rotation cover was hit by the lead shield and detached.It was not retained by the safety chain and brushed the patient as it fell.The cover was moved out of the way and the case was continued.No harm has been reported to philips.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.During the investigation philips has confirmed that the l-arm cover did not retain as the clip for the chain was bent open.The philips engineer replaced and reinstalled the l-arm cover.The system was returned to use in good working order.Philips has completed a good faith effort to collect patient information.However, the customer has not provided the requested information.Codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURION
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15425837
MDR Text Key300257693
Report Number3003768277-2022-00425
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00884838085268
UDI-Public00884838085268
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181830
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAZURION 7 M20
Device Catalogue Number722079
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received08/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/10/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age66 YR
Patient SexFemale
Patient Weight61 KG
-
-