The mayfield modified skull clamp (a1059) was not returned for evaluation after three good faith attempts (gfes) were made: device history record (dhr) review - lot number information has been provided; therefore, dhr was reviewed, and no anomalies were found.Failure analysis - based on customer provided photographs, the complaint is confirmed.There appears to be damage or wear at the opening of the starburst threads on the main clamp body.Root cause -the definite root cause cannot be identified as the device was not returned.Based on the reported complaint, probable root cause includes damage from improper assembly or mishandling.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|