• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Model Number A1059
Device Problem Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2022
Event Type  malfunction  
Event Description
A facility reported that there was an evidence of wear on the thread of the head which affects the moment of anchoring the mayfield modified skull clamp (a1059).The unit was used for a craniotomy procedure which was delayed for 20 to 30 minutes with no patient injury reported.The clinical team intervened making an inappropriate adjustment and the surgery was successfully completed without patient injury.
 
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Manufacturer Narrative
The mayfield modified skull clamp (a1059) was not returned for evaluation after three good faith attempts (gfes) were made: device history record (dhr) review - lot number information has been provided; therefore, dhr was reviewed, and no anomalies were found.Failure analysis - based on customer provided photographs, the complaint is confirmed.There appears to be damage or wear at the opening of the starburst threads on the main clamp body.Root cause -the definite root cause cannot be identified as the device was not returned.Based on the reported complaint, probable root cause includes damage from improper assembly or mishandling.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key15433562
MDR Text Key303668617
Report Number3004608878-2022-00189
Device Sequence Number1
Product Code HBL
UDI-Device IdentifierM268A10591
UDI-Public10381780253457
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA1059
Device Catalogue NumberA1059
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/02/2022
Date Device Manufactured05/03/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-