MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problems Failure to Interrogate (1332); Device Difficult to Program or Calibrate (1496); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2022
Event Type  malfunction  
Event Description
It was reported that the mobile programmer was unable to establish telemetry with the leadless implantable pulse generator (ipg) in a patient who also had a subcutaneous implantable cardioverter defibrillator (icd) and a left ventricular assist device (lvad).The user then switched to a legacy programmer and telemetry was intermittent finally losing telemetry during changing of programming with certain programming changes achieved.It was further noted that troubleshooting assistance was provided and telemetry was unable to be re-established despite the user taking recommended steps of attempting multiple angles, using a different radiofrequency head and shielding the battery pack of the lvad with lead apron.At this time the physician was satisfied to conclude with the programming changes which were able to be achieved prior to complete loss of telemetry.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key15435667
MDR Text Key300901622
Report Number2182208-2022-02962
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00613994175397
UDI-Public00613994175397
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? No
Date Manufacturer Received08/18/2022
Date Device Manufactured10/03/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
MC1VR01US IPG, 24970A PROGRAMMER
Patient Age55 YR
Patient SexMale
Patient Weight113 KG
-
-