Model Number 71938-01 |
Device Problems
Display or Visual Feedback Problem (1184); Unable to Obtain Readings (1516)
|
Patient Problems
Hypoglycemia (1912); Loss of consciousness (2418)
|
Event Date 08/15/2022 |
Event Type
Injury
|
Event Description
|
A customer reported being unable to test due to an unresponsive display of the adc device.The customer reported experiencing a loss of consciousness and was provided glucose gel by a family member for treatment.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre reader were reviewed and the dhrs showed the libre reader passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Manufacturer Narrative
|
Reader (b)(6) has been returned and investigated.Visual inspection has been performed on the reader and no issues were observed.Touchscreen test was performed and touchscreen responded properly.Therefore, issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A customer reported being unable to test due to an unresponsive display of the adc device.The customer reported experiencing a loss of consciousness and was provided glucose gel by a family member for treatment.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|