Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacturing of the device or prior repair of the device.A product sample was received for evaluation.Visual inspection and functional testing were performed.The case of the reported device was observed to be damaged at three location.The reported device was powered up with batteries and was tested on a simulator with an adult probe and pediatric probe and all readings were accurate and steady.The technician was not able to duplicate the reported issue.The disposable infant probe was returned with the reported device, and it looks like it was reused several times as wires were showing, and the probe was no longer sticky.The reported device was getting hot after running for several hours which is normal.The technician was able to confirm an off label use of the disposable/ one time use probe.No information has been received to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
|