MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2022
Event Type  malfunction  
Event Description
Information was received from a consumer and a healthcare provider (hcp) via a company representative regarding a patient receiving dilaudid (2 mg/ml at 0.35 mg/day with a 0.035 mg bolus up to 10x/day) via an implanted pump.The indication for pump use was spinal pain.The patient reported that today her pump started alarming and her ptm (personal therapy manager) was showing an alert "pump reset has occurred and is set to a minimum rate".She called her hcp who instructed her to call in.She stated that she was able to clear out of the alert but was seeing pa (patient activation) disabled.She had not recently had an mri.The patient was redirected to her hcp to further address the issue.Additional information was received on (b)(6)2022 from the hcp via a company representative who reported that the pump logs showed that the pump had a ¿reset due to firmware error¿ and had defaulted to minimum rate.It was reviewed that the hcp should first silence the active alarm and update the pump with the desired settings.The reporter was going to relay the information to the hcp.Additional information was received later the same day at which time assistance was being requested in silencing the pump alarm and reprogramming.The steps to silence the active alarm and program the pump back to the desired settings including the ptm was reviewed during the call.The pump was successfully updated and upon reinterrogation no alerts were found.The troubleshooting steps performed during the call resolved the reported issue.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15444557
MDR Text Key300251408
Report Number3004209178-2022-11983
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00763000422608
UDI-Public00763000422608
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2022
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Date Manufacturer Received09/12/2022
Date Device Manufactured06/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age43 YR
Patient SexFemale
Patient Weight54 KG
-
-