During the ablation procedure, signals could not be obtained and an impedance issue was noted resulting in procedure cancellation.It was not possible to get ablation signals on the systems.Additionally, the generator displayed an impedance out of range message with all 3 catheters.The issue was resolved by replacing the tacticath se catheter with no consequences to the patient.The signal issue with the ablation catheters was due to the gen connect cable being loose from the generator to x.However, after that issue was resolved, the generator read impedance out of range after 1-2 minutes with all 3 ablation catheters.The generator was shut down, turned back on and the impedance was normal for 1-2 minutes.After time spent troubleshooting cable issues, the patients pathway was non inducible, and the procedure was cancelled.
|