The device was not returned to zoll medical corporation; instead, the electrode pads were returned, in the original packaging.Evaluation of the electrode pads showed the packaging had minor wrinkles.The pads were stuck to the correct side of the styrene liner making it easy to remove the pads without causing any damage.Visual inspection of the pads showed no signs of arcing.The pads also passed continuity testing.No burning odor was observed during the evaluation.Review of the provided data files confirmed the event started at 11:14, on 8/25/22.The log showed a single shock via synchronized cardioversion discharged at 260j with the joule setting at 200j and the patient impedance at 233 ohms.Based off the observed metric, the shock was delivered within specification.The value of the patient impedance appears to be high and is indicative of poor coupling.There are multiple factors outside of a device malfunction that could contribute to this condition.Some include, but are not limited to motion artifacts, or poor contact between the pads and the patient's skin.This report has been attributed to poor coupling.Analysis of reports of this type has not identified an increase in trend.
|