This follow-up report is being submitted to relay additional information.A visual inspection was conducted on the provided picture of the plate cutter.The device shows signs of multiple uses including marking and scratching on the instrument surface.One of the two tips of the instrument has fractured.The complaint is confirmed.A determination cannot be made as to what caused the instrument to fracture.Review of the dhr identified no deviations or anomalies.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.This report is being submitted to update additional information in section b4, b5, g3, g6, h2, h3, h4, h6 and h10.
|