• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA PLUS3 WHITE PEGS; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA PLUS3 WHITE PEGS; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 394601
Device Problems Failure to Disconnect (2541); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/29/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone #: (b)(6).A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd connecta plus3 white pegs could not be removed.The following information was provided by the initial reporter, translated from chinese to english: after connecting the three-way with the deep vein catheter, it could not be removed.
 
Manufacturer Narrative
H6: investigation summary: our quality engineer inspected the 1 photo submitted for evaluation.The reported issue of connection issues was not confirmed but component damage - no leak was confirmed upon inspection of the photo.The photo provided shows port damaged / broken into another device.Based off the damage observed in the photo the cause of the component damage is likely improper handling of the device or use of excessive force during device implementation.It is recommended that prior to the use of bd products to review the instructions for use documentation supplied to ensure the greatest chances of there being no failures during use.Bd cannot confirm the cause of the connection issues failure to our manufacturing process since no sample was returned for evaluation.The device history records (dhr) review was performed for the lot number material identified in this complaint.According to the documented records, the product was manufactured according to the approved manufacturing procedures, specifications, and then released by quality assurance.
 
Event Description
It was reported that the bd connecta plus3 white pegs could not be removed.The following information was provided by the initial reporter, translated from chinese to english: after connecting the three-way with the deep vein catheter, it could not be removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CONNECTA PLUS3 WHITE PEGS
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15460559
MDR Text Key306327435
Report Number9610847-2022-00358
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number394601
Device Lot Number1123685
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-