|
Model Number 04.004.449S |
Device Problem
Material Deformation (2976)
|
Patient Problems
Failure of Implant (1924); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 09/03/2022 |
Event Type
Injury
|
Event Description
|
It was reported that on (b)(6) 2022, a procedure to remove a tibial nail was performed.The implant had been bent due to trauma as a result of a motorcycle accident.No further information is available.This report involves one 10mm ti cannulated tibial nail-ex/345mm-sterile.This is report 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device available for evaluation: complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter occupation: reporter is a j&j employee.(b)(4).The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: product code: 04.004.449s lot number: 272p702 manufacturing site: mezzovico release to warehouse date: 02 aug 2021 expiration date: 01 july 2031 a manufacturing record evaluation was performed for the sterile finished lot number, and no non-conformances were identified.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned photo.Visual analysis of the photo revealed that 10 ti cannulated tibial nail-ex/345-sile had bent.With the x-ray provided the allegation can be confirmed.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for 10 ti cannulated tibial nail-ex/345-sile.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|