• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION; ELECTRODE, PACEMAKER, TEMPORARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION; ELECTRODE, PACEMAKER, TEMPORARY Back to Search Results
Model Number PE075F5
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2022
Event Type  malfunction  
Manufacturer Narrative
The reported event of balloon inflation issue was confirmed.Balloon did not inflate due to leakage from a tear at distal side of balloon latex.After cutting the proximal side of balloon, it was confirmed the balloon latex edges were not able to match up.No visible damage was found from catheter body, returned syringe and balloon windings.Further evaluation regarding related quality issues is under investigation.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event Description
It was reported that the balloon of a swan ganz catheter could not maintain the inflation during balloon test before use.There were no patient complications reported.Per product evaluation, balloon was found ruptured with ruptured edges not matching.
 
Manufacturer Narrative
Engineering evaluation was completed and concluded that a product risk assessment escalation was met.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine 92614
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
MDR Report Key15476782
MDR Text Key305883366
Report Number2015691-2022-08070
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date12/06/2023
Device Model NumberPE075F5
Device Lot Number64088336
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-