MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION EVOLUT FX VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION EVOLUT FX VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number EVOLUTFX-29
Device Problems Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Patient Problem Vascular Dissection (3160)
Event Date 09/08/2022
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products:other relevant device(s) are: product id: d- evolutfx-2329, serial/lot #: (b)(4), ubd: 21-jun-2023, udi#: (b)(4).Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during the implant of this transcatheter bioprosthetic valve into a large patient with a steep implant angles (caudal of 37 degrees), the valve was deployed in a position of 0-1 millimeter (mm) on the non-coronary cusp (ncc).The valve was released and upon removing the delivery catheter system (dcs), it appeared the valve was no longer in the annulus.A snare was used to capture the tab on the inner curve and was pulled up above the sinotubular junction and held in place.A new valve was loaded onto a new dcs.Multiple attempts were made to cross the annulus was unsuccessful.The non-medtronic guidewire was exchanged for another non-medtronic guidewire.Upon removal of the dcs, the flare capsule by the nosecone was flared out, most likely caused by catching the first valve.A new valve was prepped.An angiogram revealed an ascending aortic dissection.The guidewire was changed and the valve was implanted into the annulus.Another angiogram was performed and the physician believes the valve was holding the dissection flap closed.Per the physician, the valve was implanted higher than what they thought.The patient is doing well and stable.Per the physician, there was no issue with the valve or dcs however was due to body habitus, poor imaging quality, and high implant.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Conclusion: difficulties advancing the delivery catheter system (dcs) through the access vessel is known to be related to factors such as patient anatomy and physician technique, including guidewire and introducer sheath selection.In this case, it was noted that the patient had a steep implant angle.This indicates that the probable cause of the advancement difficulties was patient anatomy, but this cannot be confirmed with the limited information available.Advancement difficulties do not typically indicate a device malfunction or a failure to meet manufacturing specifications.Potential factors that can influence a dislodged valve include tension applied on the dcs during positioning, calcification levels in the native vessel, compliance of the aorta and native vessels, and a number of others, but in this case, dcs interaction caused the original valve to dislodge.In addition, the valve may have been undersized, which could have contributed to the dislodgement.Dislodge events are typically not related to a device malfunction.Vascular injuries such as dissection, are known potential adverse patient effect per the ifu, and are typically related to patient factors (anatomy, comorbidities, etc.), and/or procedural effects (sheath used, user technique, puncture cut location, etc.).It was reported that, per the physician, the snaring of the valve caused the aortic dissection.The physician believed the valve was holding the dissection flap closed.Repositioning or snaring the bioprosthesis is not recommended, except in cases where imminent serious harm or death is possible.Repositioning of a deployed valve may cause aortic root damage, coronary artery damage, myocardial damage, vascular complications, prosthetic valve dysfunction among others.The device history record for the dcs was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.There is no information to suggest a device quality deficiency that may have caused or contributed to this event.Updated: h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information was received which indicated that the first valve did not dislodge due to the delivery catheter system (dcs) removal.As reported, the second dcs did not cause the dissection.The cause was not reported.The dissection did not require treatment.No additional adverse patient effects were reported.Update data: annex g code.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information was received that the deployment starting point was mid-pigtail and the valve dislodged aortic.The implant depth on the left coronary cusp (lcc) was 2-3 mm at the time of release.Per the physician, the cause of the dissection was valve snaring.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT FX VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15478184
MDR Text Key300579876
Report Number2025587-2022-02637
Device Sequence Number1
Product Code NPT
UDI-Device Identifier00763000370701
UDI-Public00763000370701
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/12/2023
Device Model NumberEVOLUTFX-29
Device Catalogue NumberEVOLUTFX-29
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2022
Date Device Manufactured09/12/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age78 YR
Patient SexMale
Patient Weight135 KG
-
-