MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN STRAIGHT CUP IMPACTOR; HIP INSTRUMENTS : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN STRAIGHT CUP IMPACTOR; HIP INSTRUMENTS : IMPACTORS Back to Search Results
Model Number 2217-50-041
Device Problems Improper or Incorrect Procedure or Method (2017); Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2022
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the thread on tip was damaged.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Additional information received: a.Can you please clarify what do you mean by ¿damaged¿? was it worn, cracked, broken into pieces, bent, stripped, cross threaded or any device interaction.The thread is damaged b.What was the duration of delay? no delay.C.Please verify if there was any alleged device deficiency reported for the hardinge cement restrictor as also mentioned in the email trial in the attached documents? if yes, was this already reported? please provide pc number? no defective for hardinge cement restrictior, i just needed another one because one was used in a emergency situation at the private hospital and i borrowed the one from (b)(6).
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The investigation could not verify or draw any conclusions about the root cause of the reported event without the device to examine.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary the device associated with this report was returned to depuy synthes for evaluation.Visual examination of the device revealed that the threaded insert area is stripped, therefore the allegation is confirmed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthese joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN STRAIGHT CUP IMPACTOR
Type of Device
HIP INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15490275
MDR Text Key301083890
Report Number1818910-2022-18773
Device Sequence Number1
Product Code HWA
UDI-Device Identifier10603295098980
UDI-Public10603295098980
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2217-50-041
Device Catalogue Number221750041
Device Lot NumberSO2050630
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/14/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-