ST PAUL PORTEX EPIDURAL CONTINUOUS TRAYS; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 30-4498-001
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2020
Event Type  malfunction  
Event Description
It was reported that when inserting the catheter into the epifuse connector during the use of the product, the customer noticed the connector broke off.No patient injury was reported.
 
Manufacturer Narrative
No lot number was provided; therefore, device history record review could not be completed.A product sample was received for evaluation.Visual and functional testing were performed.As a result of observing the connector, it was confirmed that the hinge was broken.Therefore, it seems that the hinge part cracked and damaged when using the connector.However, the cause and timing of occurrence were not identified.The reported event was confirmed.The root cause was not identified.Udi # is unknown.No product information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX EPIDURAL CONTINUOUS TRAYS
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
akasaka, minato-ku, tokyo
minneapolis, MN 55442
MDR Report Key15492013
MDR Text Key306092588
Report Number3012307300-2022-20354
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K965017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number30-4498-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/13/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-