• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION ASSY,MAIH,2.0 DIA, W/SWITCH, 100-120VONLY, PACKAG; INTERNAL PADDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION ASSY,MAIH,2.0 DIA, W/SWITCH, 100-120VONLY, PACKAG; INTERNAL PADDLE Back to Search Results
Model Number 1011-0139-05
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that during biomed testing, the associated defibrillator was unable to charge using these attached internal handles.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
The internal handles will not be returned to zoll medical corporation.The customer reported to zoll medical corporation that these internal handles were manufactured in 2014 and have gone through over 100 sterilization cycles.Due to the age of these internal handles (8 years), they have most likely met their useful life and have exceeded the maximum of 50 autoclave/sterilization cycles.The customer will remove this set of internal handles from service and will order a replacement set.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSY,MAIH,2.0 DIA, W/SWITCH, 100-120VONLY, PACKAG
Type of Device
INTERNAL PADDLE
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key15496834
MDR Text Key300714866
Report Number1220908-2022-03655
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1011-0139-05
Device Catalogue Number8011-0139-05
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-