ZOLL MEDICAL CORPORATION ASSY,MAIH,2.0 DIA, W/SWITCH, 100-120VONLY, PACKAG; INTERNAL PADDLE
|
Back to Search Results |
|
Model Number 1011-0139-05 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during biomed testing, the associated defibrillator was unable to charge using these attached internal handles.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The internal handles will not be returned to zoll medical corporation.The customer reported to zoll medical corporation that these internal handles were manufactured in 2014 and have gone through over 100 sterilization cycles.Due to the age of these internal handles (8 years), they have most likely met their useful life and have exceeded the maximum of 50 autoclave/sterilization cycles.The customer will remove this set of internal handles from service and will order a replacement set.Analysis of reports of this type has not identified an increase in trend.
|
|