Model Number 2544-00-525
Device Problems Material Deformation (2976); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2022
Event Type  malfunction  
Manufacturer Narrative
Product complaint #(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that femoral sizer is sticking and will not move up and down freely.The central metal rod was bent.There was no surgical delay.
 
Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was returned.After physical review, the allegation cannot be confirmed due to the deformed/bent condition reported was not observed in the device.Additionally, the jammed allegation was not identified due to the sliding platform works as intended.No issues were identified.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mrae) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE MES SIZING/ROT GDE
Type of Device
ATTUNE INSTRUMENTS : SIZING/MEASURING INSTRUMENTS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15501495
MDR Text Key305710071
Report Number1818910-2022-18994
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295434061
UDI-Public10603295434061
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2544-00-525
Device Catalogue Number254400525
Device Lot NumberAB5038224
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
-
-